Power standard based on a high precision power-factor meter
- Autor(es):
- Slomovitz, Daniel ; Faverio, Carlos
- Tipo:
- Artículo
- Versión:
- Publicado
- Resumen:
-
A high-precision power-factor meter, based on zero crossing detectors, is proposed. The influence of the distortion in the input signals is analyzed and low pass filters proposed to prevent this error source. This system is proposed as a new power standard meter, based on the measurement of voltage, current and power factor.
- Año:
- 1998
- Idioma:
- Inglés
- Temas:
- Distortion measurement
Voltage
Current measurement
Instruments
Harmonic distortion
Power measurement
Reactive power
Laboratories
Detectors
Voltmeters
- Institución:
- Universidad de la República
- Repositorio:
- COLIBRI
- Enlace(s):
- https://hdl.handle.net/20.500.12008/20765
- Nivel de acceso:
- Acceso abierto