Sample temperature affects measurement of nitrate with a rapid analysis ion selective electrode system used for N management of vegetable crops.
Resumen:
ABSTRACT.- The practical value of portable hand-held ion selective electrode sensors (ISE) for on-farm [NO3-] measurement to assist with crop N management of vegetable crops has been demonstrated in numerous previous studies. They provide rapid, in-situ measurement of the nitrate concentration ([NO3-]) in nutrient and soil solutions, and in petiole sap. Sample temperatures, for on-farm measurements, vary appreciably. This study evaluated the effects of sample temperature on [NO3-] measurement using two different models of a commonly used, commercially available, portable ISE meter. The temperatures (5, 10, 15, 20, and 25 °C) examined were in the range likely to be encountered in practical on-farm work. Aqueous solutions of 6, 12, and 18 mmol NO3-L-1 were prepared from KNO3, Ca(NO3)2 and NaNO3. [NO3-] was measured in three replicate samples of each of the three concentrations, made from each NO3-; compound, at each temperature. The results consistently and clearly demonstrated a strong negative linear relationship between temperature-induced errors and sample temperatures. The temperature-induced error was considerable for cooled samples, being +50% at 5 °C and +31% at 10 °C. At sample temperatures of 17-20 °C, the temperature effects were minimal. Above this range, the temperature effect caused underestimation. At 25 °C, the temperature-induced error was -24%. These results show that care must be taken to ensure that sample temperatures do not erroneously affect the measurement of [NO3-] with ISE meters. Particular care needs to be taken with both refrigerated and warmer samples. © 2022 by the authors.
| 2022 | |
|
Quick tests ISE LAQUAtwin Horiba N monitoring Nutrient solution Soil solution Petiole sap Fertigation |
|
| Inglés | |
| Instituto Nacional de Investigación Agropecuaria | |
| AINFO | |
| https://ainfo.inia.uy/consulta/busca?b=pc&id=63825&biblioteca=vazio&busca=63825&qFacets=63825 | |
| Acceso abierto |
| _version_ | 1856772322855223296 |
|---|---|
| author | PEÑA-FLEITAS, M.T. |
| author2 | GRASSO, R. GALLARDO, M. PADILLA, F.M. DE SOUZA, R. RODRÍGUEZ , A. THOMPSON, R.B. |
| author2_role | author author author author author author |
| author_facet | PEÑA-FLEITAS, M.T. GRASSO, R. GALLARDO, M. PADILLA, F.M. DE SOUZA, R. RODRÍGUEZ , A. THOMPSON, R.B. |
| author_role | author |
| bitstream.checksum.fl_str_mv | 33c6231dcd0a4b1d101847240cf91388 |
| bitstream.checksumAlgorithm.fl_str_mv | MD5 |
| bitstream.url.fl_str_mv | https://redi.anii.org.uy/jspui/bitstream/20.500.12381/4264/1/sword-2025-06-23T15%3a25%3a46.original.xml |
| collection | AINFO |
| dc.creator.none.fl_str_mv | PEÑA-FLEITAS, M.T. GRASSO, R. GALLARDO, M. PADILLA, F.M. DE SOUZA, R. RODRÍGUEZ , A. THOMPSON, R.B. |
| dc.date.accessioned.none.fl_str_mv | 2025-06-23T18:25:46Z |
| dc.date.available.none.fl_str_mv | 2025-06-23T18:25:46Z |
| dc.date.issued.none.fl_str_mv | 2022 |
| dc.date.updated.none.fl_str_mv | 2025-06-23T18:25:46Z |
| dc.description.abstract.none.fl_txt_mv | ABSTRACT.- The practical value of portable hand-held ion selective electrode sensors (ISE) for on-farm [NO3-] measurement to assist with crop N management of vegetable crops has been demonstrated in numerous previous studies. They provide rapid, in-situ measurement of the nitrate concentration ([NO3-]) in nutrient and soil solutions, and in petiole sap. Sample temperatures, for on-farm measurements, vary appreciably. This study evaluated the effects of sample temperature on [NO3-] measurement using two different models of a commonly used, commercially available, portable ISE meter. The temperatures (5, 10, 15, 20, and 25 °C) examined were in the range likely to be encountered in practical on-farm work. Aqueous solutions of 6, 12, and 18 mmol NO3-L-1 were prepared from KNO3, Ca(NO3)2 and NaNO3. [NO3-] was measured in three replicate samples of each of the three concentrations, made from each NO3-; compound, at each temperature. The results consistently and clearly demonstrated a strong negative linear relationship between temperature-induced errors and sample temperatures. The temperature-induced error was considerable for cooled samples, being +50% at 5 °C and +31% at 10 °C. At sample temperatures of 17-20 °C, the temperature effects were minimal. Above this range, the temperature effect caused underestimation. At 25 °C, the temperature-induced error was -24%. These results show that care must be taken to ensure that sample temperatures do not erroneously affect the measurement of [NO3-] with ISE meters. Particular care needs to be taken with both refrigerated and warmer samples. © 2022 by the authors. |
| dc.identifier.none.fl_str_mv | https://ainfo.inia.uy/consulta/busca?b=pc&id=63825&biblioteca=vazio&busca=63825&qFacets=63825 |
| dc.language.iso.none.fl_str_mv | en eng |
| dc.rights.es.fl_str_mv | Acceso abierto |
| dc.rights.none.fl_str_mv | info:eu-repo/semantics/openAccess |
| dc.source.none.fl_str_mv | reponame:AINFO instname:Instituto Nacional de Investigación Agropecuaria instacron:Instituto Nacional de Investigación Agropecuaria |
| dc.subject.none.fl_str_mv | Quick tests ISE LAQUAtwin Horiba N monitoring Nutrient solution Soil solution Petiole sap Fertigation |
| dc.title.none.fl_str_mv | Sample temperature affects measurement of nitrate with a rapid analysis ion selective electrode system used for N management of vegetable crops. |
| dc.type.none.fl_str_mv | Article PublishedVersion info:eu-repo/semantics/article |
| dc.type.version.none.fl_str_mv | info:eu-repo/semantics/publishedVersion |
| description | ABSTRACT.- The practical value of portable hand-held ion selective electrode sensors (ISE) for on-farm [NO3-] measurement to assist with crop N management of vegetable crops has been demonstrated in numerous previous studies. They provide rapid, in-situ measurement of the nitrate concentration ([NO3-]) in nutrient and soil solutions, and in petiole sap. Sample temperatures, for on-farm measurements, vary appreciably. This study evaluated the effects of sample temperature on [NO3-] measurement using two different models of a commonly used, commercially available, portable ISE meter. The temperatures (5, 10, 15, 20, and 25 °C) examined were in the range likely to be encountered in practical on-farm work. Aqueous solutions of 6, 12, and 18 mmol NO3-L-1 were prepared from KNO3, Ca(NO3)2 and NaNO3. [NO3-] was measured in three replicate samples of each of the three concentrations, made from each NO3-; compound, at each temperature. The results consistently and clearly demonstrated a strong negative linear relationship between temperature-induced errors and sample temperatures. The temperature-induced error was considerable for cooled samples, being +50% at 5 °C and +31% at 10 °C. At sample temperatures of 17-20 °C, the temperature effects were minimal. Above this range, the temperature effect caused underestimation. At 25 °C, the temperature-induced error was -24%. These results show that care must be taken to ensure that sample temperatures do not erroneously affect the measurement of [NO3-] with ISE meters. Particular care needs to be taken with both refrigerated and warmer samples. © 2022 by the authors. |
| eu_rights_str_mv | openAccess |
| format | article |
| id | INIAOAI_9b96702eeb49240c20297158f8564aea |
| instacron_str | Instituto Nacional de Investigación Agropecuaria |
| institution | Instituto Nacional de Investigación Agropecuaria |
| instname_str | Instituto Nacional de Investigación Agropecuaria |
| language | eng |
| language_invalid_str_mv | en |
| network_acronym_str | INIAOAI |
| network_name_str | AINFO |
| oai_identifier_str | oai:redi.anii.org.uy:20.500.12381/4264 |
| publishDate | 2022 |
| reponame_str | AINFO |
| repository.mail.fl_str_mv | lorrego@inia.org.uy |
| repository.name.fl_str_mv | AINFO - Instituto Nacional de Investigación Agropecuaria |
| repository_id_str | |
| rights_invalid_str_mv | Acceso abierto |
| spelling | 2025-06-23T18:25:46Z2025-06-23T18:25:46Z20222025-06-23T18:25:46Zhttps://ainfo.inia.uy/consulta/busca?b=pc&id=63825&biblioteca=vazio&busca=63825&qFacets=63825ABSTRACT.- The practical value of portable hand-held ion selective electrode sensors (ISE) for on-farm [NO3-] measurement to assist with crop N management of vegetable crops has been demonstrated in numerous previous studies. They provide rapid, in-situ measurement of the nitrate concentration ([NO3-]) in nutrient and soil solutions, and in petiole sap. Sample temperatures, for on-farm measurements, vary appreciably. This study evaluated the effects of sample temperature on [NO3-] measurement using two different models of a commonly used, commercially available, portable ISE meter. The temperatures (5, 10, 15, 20, and 25 °C) examined were in the range likely to be encountered in practical on-farm work. Aqueous solutions of 6, 12, and 18 mmol NO3-L-1 were prepared from KNO3, Ca(NO3)2 and NaNO3. [NO3-] was measured in three replicate samples of each of the three concentrations, made from each NO3-; compound, at each temperature. The results consistently and clearly demonstrated a strong negative linear relationship between temperature-induced errors and sample temperatures. The temperature-induced error was considerable for cooled samples, being +50% at 5 °C and +31% at 10 °C. At sample temperatures of 17-20 °C, the temperature effects were minimal. Above this range, the temperature effect caused underestimation. At 25 °C, the temperature-induced error was -24%. These results show that care must be taken to ensure that sample temperatures do not erroneously affect the measurement of [NO3-] with ISE meters. Particular care needs to be taken with both refrigerated and warmer samples. © 2022 by the authors.https://hdl.handle.net/20.500.12381/4264enenginfo:eu-repo/semantics/openAccessAcceso abiertoQuick testsISELAQUAtwinHoribaN monitoringNutrient solutionSoil solutionPetiole sapFertigationSample temperature affects measurement of nitrate with a rapid analysis ion selective electrode system used for N management of vegetable crops.ArticlePublishedVersioninfo:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionreponame:AINFOinstname:Instituto Nacional de Investigación Agropecuariainstacron:Instituto Nacional de Investigación AgropecuariaPEÑA-FLEITAS, M.T.GRASSO, R.GALLARDO, M.PADILLA, F.M.DE SOUZA, R.RODRÍGUEZ , A.THOMPSON, R.B.SWORDsword-2025-06-23T15:25:46.original.xmlOriginal SWORD entry documentapplication/octet-stream3136https://redi.anii.org.uy/jspui/bitstream/20.500.12381/4264/1/sword-2025-06-23T15%3a25%3a46.original.xml33c6231dcd0a4b1d101847240cf91388MD5120.500.12381/42642026-02-10 15:53:56.579oai:redi.anii.org.uy:20.500.12381/4264Institucionalhttps://ainfo.inia.uy/Organismo científico-tecnológicohttp://inia.uyhttps://redi.anii.org.uy/oai/requestlorrego@inia.org.uyUruguayopendoar:2026-02-10T18:53:56AINFO - Instituto Nacional de Investigación Agropecuariafalse |
| spellingShingle | Sample temperature affects measurement of nitrate with a rapid analysis ion selective electrode system used for N management of vegetable crops. PEÑA-FLEITAS, M.T. Quick tests ISE LAQUAtwin Horiba N monitoring Nutrient solution Soil solution Petiole sap Fertigation |
| status_str | publishedVersion |
| title | Sample temperature affects measurement of nitrate with a rapid analysis ion selective electrode system used for N management of vegetable crops. |
| title_full | Sample temperature affects measurement of nitrate with a rapid analysis ion selective electrode system used for N management of vegetable crops. |
| title_fullStr | Sample temperature affects measurement of nitrate with a rapid analysis ion selective electrode system used for N management of vegetable crops. |
| title_full_unstemmed | Sample temperature affects measurement of nitrate with a rapid analysis ion selective electrode system used for N management of vegetable crops. |
| title_short | Sample temperature affects measurement of nitrate with a rapid analysis ion selective electrode system used for N management of vegetable crops. |
| title_sort | Sample temperature affects measurement of nitrate with a rapid analysis ion selective electrode system used for N management of vegetable crops. |
| topic | Quick tests ISE LAQUAtwin Horiba N monitoring Nutrient solution Soil solution Petiole sap Fertigation |
| url | https://ainfo.inia.uy/consulta/busca?b=pc&id=63825&biblioteca=vazio&busca=63825&qFacets=63825 |