Wearable estimation of central aortic blood pressure.
Supervisor(es): Silveira, Fernando - Armentano, Ricardo
Resumen:
Arterial hypertension affects a third of the world's population and is a significant risk factor for cardiovascular disease. Blood pressure (BP) is one of the most relevant parameters used for monitoring of possible hypertension states in patients at risk of cardiovascular disease. Hence, there exists a need for new monitoring solutions, which allow to increase the frequency between BP assessments, but also allow to reduce the level of occlusion in the attempts. Moens-Korteweg equation is among the main principles to estimate BP by dispensing of any inflatable cuff. This principle might lead to an indirect estimation of BP by measuring the time it takes the pressure pulse to propagate between two pre-established vascular points, accordingly the pulse transit time (PTT) method. This thesis proposes a wearable PTT-based method to estimate central aortic BP (CABP) and, the main milestones of this work included: proof of concept of the proposed method (pilot work), the development of a wearable device (including two stages of validation), the proposition of a miniaturized version (integrated circuit) of the analog front-end of the wearable hardware, and, the development of a novel PTT-based model (PTTBM, i.e., the mathematical relationship between measured variables and estimated BP) suitable for the proposed wearable methodology to estimate BP. The main contributions found at each milestone are presented. One of the contributions of this thesis is the use of the PTT-principle for estimating CABP instead of the peripheral BP (PBP) (as typically used in the literature). The pilot work showed the feasibility of CABP estimation from the PTT principle by using electrocardiogram (ECG) and ballistocardiogram (BCG) recordings from off-the-shelf equipment. Results showed that CABP was more correlated with the proposed methodology in comparison to all PBP variables assessed; confirming our hypothesis that the CABP is the most suitable parameter to collate through the time elapsed from ECG R-wave to the BCG J-wave. That is, considered featured time (RJ-interval) includes the time of a pulse pressure propagating at an aortic district. Bland-Altman plots showed an almost zero mean error (\u\ < 0.02mmHg) and bounded standard deviation o < 5mmHg for all systolic and mean central BP readings. Pilot work provided a landmark in order to develop a compact device that allows the integration of wireless blood pressure monitoring into a wearable system. Another contribution of this thesis is the proposition of a wearable device for PTT-computing by also including design considerations for the signal conditioning chains for ECG and BCG signals. The proposed design procedure takes care of minimizing the impact of spurious delays between physiological signals, which eventually degrade the PTT computation. Further, such a procedure could be suitable for any PTT-acquisition. Filtering with low and controlled delay is required for this biomedical application, and proposed conditioning chains provide less than 2ms group-delay, showing the effectiveness of the proposed approach. In order to provide the methodology with higher autonomy and integration, a highly miniaturized implementation of the filtering approach was also proposed. It includes the design of proposed architectures in CMOS technology to implement the particular low-delay filtering at reduced bandwidth featuring ultra-low-power characteristics. Results show that less than 2ms delay for the ECG QRS-complex can be achieved with a total current consumption of IDD = 2:1nA at VDD = 1:2V of power supply. Such development meant another significant contribution of this work in the conception of highly autonomous wearable devices for PTT acquisition. The first stage of validations on the wearable CABP estimation showed that, when considering data from one volunteer, results achieved with off-the-shelf equipment could be replicated by using a proposed wearable device, and the method could be further validated by using the wearable version. Additionally, CABP estimation from the proposed wearable device could be feasible by using three feature times (FTs) as CABP surrogates; that is, RI, RJ, and IJ intervals (from ECG and BCG wearable recordings). The first validation of the method also showed that CABP could be accurately predicted by the proposed methodology when in the order of daily calibrations are performed. The second stage of validations involved a study with a group of volunteers, and new alternatives were explored (twentyseven: nine PTTBMs along the three FTs) for the CABP estimation. We found that CABP could be accurately estimated (inside AAMI requirements) through the presented methodology by using four of the explored alternatives, whereas the RI interval, an FT lacking any PTT assessment, emerged as the best surrogate for the CABP estimation. Hence, a principle different from the traditional PTT-based method arises as a more advantageous method for the CABP estimation in the light of evidence reported in this validation, and, to our knowledge, this is the first time that CABP has been successfully estimated from a wearable device. The final significant contribution of this thesis meant the last chain-link in the process to achieve an utterly original method to estimate CABP. A novel PTTBM to estimate CABP is proposed, which uses a ow-driven two-element Windkesel network constructed from FTs extracted from the wearable recordings. When classic PTTBMs are applied, the fitting of parameters often leads to values without a physiological basis. Opposite to that in the proposed PTTBM, the parameters have a clear physiological meaning, and the parameter fitting led to values that are consistent with this meaning and more stable throughout calibrations. In conclusion, this thesis introduces a novel device that exploits an alternative and indirect method for CABP estimation. Variants of the principle used, accordingly, PTT method, have been previously explored to estimate PBP but not for central aortic BP. Additionally, the device was designed to be wearable; that is, it is attached to the clothes, causing low discomfort for the user during the measurement, thus, allowing continuous and ambulatory monitoring of aortic pressure. The developed wearable system, validated in a series of volunteers, showed promising results towards the continuous CABP monitoring.
Se estima que casi un tercio de la población adulta mundial sufre de algún grado de hipertensión, siendo esto un factor de riesgo significativo para la enfermedad cardiovascular. La presión arterial (PA) es el parámetro utilizado para evaluar estos posibles estados de hipertensión; actualmente existe una necesidad de generación de nuevas tecnologías que permitan aumentar la frecuencia entre medidas de PA, pero al mismo tiempo de reducir el nivel de oclusión de éstas (técnicas aceptadas están mayoritariamente basadas en la oclusión y son de acceso limitado). El modelo Moens-Korteweg podría proveer los argumentos para la creación de nuevas técnicas para estimar la PA prescindiendo de cualquier brazalete inflable. Más específicamente, podría obtenerse una estimación indirecta de la PA a través de la medición del tiempo que tarda el pulso de presión en propagarse entre dos puntos vasculares predefinidos, método conocido como tiempo de tránsito del pulso (PTT). En la presente tesis se desarrolló un dispositivo vestible que explota este método alternativo e indirecto para la estimación de la PA pero a nivel central, es decir, busca estimar la PA en la aorta (CABP), la principal arteria de la red vascular. Para ello, los principales desarrollos de este trabajo incluyeron : prueba de concepto del método propuesto basado en PTT para estimar CABP, el desarrollo de un dispositivo vestible (incluyendo dos etapas de validaciones para la estimación de la PA), la propuesta de un circuito integrado para el hardware vestible y el desarrollo de un nuevo modelo para la estimación de la PA (PTTBM, es decir, la relación matemática que vincula las variables medidas con el hardware diseñado y la estimación de la PA). A continuación se presentan las principales contribuciones resultantes de cada frente de trabajo. Una de las contribuciones de esta tesis es el uso del principio PTT para estimar CABP en lugar de la BP periférica (PBP) (como se usa típicamente en la literatura). La prueba de concepto mostró la viabilidad de la estimación de CABP a partir del principio PTT mediante la adquisición de señales electrocardiograma (ECG) y balistocardiograma (BCG) utilizando equipos comerciales. Los resultados mostraron que CABP estaba más correlacionado con la metodología propuesta en comparación con todas las variables de PBP evaluadas; confirmando nuestra hipótesis de que la CABP sería la variable más adecuada para estimar a partir del tiempo transcurrido desde la onda R del ECG hasta la onda J del BCG. Es decir, el tiempo considerado (intervalo RJ) incluye un tiempo de propagación del pulso de presión a través de un segmento aórtico. Las gráficas de Bland-Altman mostraron un error medio casi nulo (\u\ < 0.02mmHg) y una precisión o < 5mmHg para las variables de presión sistólica y media centrales. La prueba de concepto proporcionó un hito para desarrollar un dispositivo vestible apuntando a la monitorización inalámbrica de la presión arterial en un sistema imperceptible para el usuario. Otra contribución de esta tesis es la propuesta de este dispositivo vestible para la adquisición de la PTT. El desarrollo incluye consideraciones de instrumentación necesarias para el correcto acondicionamiento de las señales ECG y BCG, de las cuales se obtiene la PTT. En particular, el procedimiento de diseño propuesto busca minimizar el impacto de los retrasos espurios entre las señales fisiológicas, que eventualmente degradan la computación de la PTT. Además, dicho procedimiento podría ser aprovechado por otros desarrolladores del método sin importar las definiciones de PTT que éstos usen. La limitación de banda con bajo retardo es necesario para esta aplicación biomédica, y el hardware de acondicionamiento propuesto proporciona menos de 2 ms de retraso en las se~nales (ECG y BCG) mientras consigue limitar sus bandas a decenas de Hz, lo que muestra la efectividad de la metodología propuesta. Adicionalmente, con el fin de proporcionar a la metodología de una mayor autonomía e integración, se propone una implementación altamente miniaturizada de la sección de filtrado con bajo retraso. Se incluye el diseño de nuevas topologías propuestas en tecnología CMOS para implementar el particular filtro de bajo retraso con reducido ancho de banda, y con características de ultra bajo consumo de potencia. El diseño integrado consigue obtener resultados similares al obtenido anteriormente (con componentes discretos) alcanzando un retraso de menos de 2 ms para el complejo QRS del ECG, pero con un consumo de IDD = 2:1 nA a un VDD = 1:2 V . Tal desarrollo significó otra contribución de este trabajo en el área de circuitos altamente autónomos para instrumentación biomédica. La primera etapa de validaciones en la estimación vestible de la CABP se basó en experimentaciones con un voluntario, mostrando que, la estimación vestible podría alcanzar los mismos resultados que los alcanzados utilizando equipos de investigación, permitiendo así avanzar en la validación del método propuesto utilizando el equipamiento vestible diseñado. Además de esto, se encontró que la estimación de CABP a partir del dispositivo vestible podría ser factible utilizando varios tiempos característicos (FT) extraídos de las señales vestibles ECG y BCG (intervalos RI, RJ e IJ) junto con un popular PTTBM. La primera validación del método también arrojó que la metodología propuesta podría estimar con precisión la CABP cuando el tiempo entre calibraciones es del orden de un día. La segunda etapa de validación implicó un estudio con un grupo de voluntarios, nuevas alternativas se exploraron esta vez (veintisiete: nueve PTTBM con tres FT) para la estimación de CABP. Descubrimos que CABP podría estimarse con precisión (dentro de los requisitos de AAMI) a través de la metodología presentada mediante el uso de cuatro de las alternativas exploradas, mientras que el intervalo RI, siendo un FT que a priori no tiene ninguna vinculación con un PTT, surge como el mejor estimador de la CABP. Se concluye entonces, que un principio diferente del método tradicional basado en PTT podría ser más ventajoso para la estimación de CABP a la luz de la evidencia encontrada en esta validación y, adicionalmente, a nuestro entender, esta es la primera vez que CABP se estima con éxito a partir de un dispositivo vestible. La contribución final de esta tesis significó el último eslabón de la cadena en el proceso de lograr un método completamente original para estimar CABP de punta a punta. Se propone un nuevo PTTBM para estimar CABP, éste es basado en una red Windkesel de dos elementos bajo una excitación de flujo. Estos elementos del PTTBM son construidos a partir de cantidades extraídas a través de procesamiento de las señales vestibles ECG y BCG. Cuando se aplican los PTTBM clásicos, el ajuste de sus parámetros (en calibración) a menudo conducen a valores sin base fisiológica, mostrando a su vez, una dispersión en sus valores a lo largo de distintas calibraciones que podrían ser inaceptables en la práctica. En contraposición, los parámetros del PTTBM propuesto convergen a cantidades con significado fisiologico claro y estable a lo largo de las calibraciones. En conclusión, esta tesis presenta un dispositivo novedoso que explota un método alternativo e indirecto para la estimación de CABP. El método propuesto es basado en la metodología de PTT, que si bien ha sido previamente explotado para estimar PBP, no se ha dirigido éste hacia el monitoreo vestible de la PA aórtica central. En este marco se desarrolla un dispositivo vestible, causando baja molestia en el usuario durante las mediciones, lo que permitiría un monitoreo continuo y ambulatorio real de la presión aórtica central. El sistema vestible desarrollado, validado en una serie de voluntarios, ha mostrado resultados prometedores hacia el monitoreo continuo de CABP.
2020 | |
Wearable devices Blood pressure instrumentation modeling |
|
Inglés | |
Universidad de la República | |
COLIBRI | |
https://hdl.handle.net/20.500.12008/25156 | |
Acceso abierto | |
Licencia Creative Commons Atribución - No Comercial - Sin Derivadas (CC - By-NC-ND 4.0) |
_version_ | 1807523178622222336 |
---|---|
author | Fierro, Germán |
author_facet | Fierro, Germán |
author_role | author |
bitstream.checksum.fl_str_mv | 6429389a7df7277b72b7924fdc7d47a9 a006180e3f5b2ad0b88185d14284c0e0 d77747f0b79dbc4c411d2260a3d95cd2 1996b8461bc290aef6a27d78c67b6b52 ec0cba14edfb7a3a945ca205978223d5 |
bitstream.checksumAlgorithm.fl_str_mv | MD5 MD5 MD5 MD5 MD5 |
bitstream.url.fl_str_mv | http://localhost:8080/xmlui/bitstream/20.500.12008/25156/5/license.txt http://localhost:8080/xmlui/bitstream/20.500.12008/25156/2/license_url http://localhost:8080/xmlui/bitstream/20.500.12008/25156/3/license_text http://localhost:8080/xmlui/bitstream/20.500.12008/25156/4/license_rdf http://localhost:8080/xmlui/bitstream/20.500.12008/25156/1/Fie20.pdf |
collection | COLIBRI |
dc.contributor.filiacion.none.fl_str_mv | Fierro Germán, Universidad de la República (Uruguay). Facultad de Ingeniería. |
dc.creator.advisor.none.fl_str_mv | Silveira, Fernando Armentano, Ricardo |
dc.creator.none.fl_str_mv | Fierro, Germán |
dc.date.accessioned.none.fl_str_mv | 2020-09-04T15:12:12Z |
dc.date.available.none.fl_str_mv | 2020-09-04T15:12:12Z |
dc.date.issued.none.fl_str_mv | 2020 |
dc.description.abstract.none.fl_txt_mv | Arterial hypertension affects a third of the world's population and is a significant risk factor for cardiovascular disease. Blood pressure (BP) is one of the most relevant parameters used for monitoring of possible hypertension states in patients at risk of cardiovascular disease. Hence, there exists a need for new monitoring solutions, which allow to increase the frequency between BP assessments, but also allow to reduce the level of occlusion in the attempts. Moens-Korteweg equation is among the main principles to estimate BP by dispensing of any inflatable cuff. This principle might lead to an indirect estimation of BP by measuring the time it takes the pressure pulse to propagate between two pre-established vascular points, accordingly the pulse transit time (PTT) method. This thesis proposes a wearable PTT-based method to estimate central aortic BP (CABP) and, the main milestones of this work included: proof of concept of the proposed method (pilot work), the development of a wearable device (including two stages of validation), the proposition of a miniaturized version (integrated circuit) of the analog front-end of the wearable hardware, and, the development of a novel PTT-based model (PTTBM, i.e., the mathematical relationship between measured variables and estimated BP) suitable for the proposed wearable methodology to estimate BP. The main contributions found at each milestone are presented. One of the contributions of this thesis is the use of the PTT-principle for estimating CABP instead of the peripheral BP (PBP) (as typically used in the literature). The pilot work showed the feasibility of CABP estimation from the PTT principle by using electrocardiogram (ECG) and ballistocardiogram (BCG) recordings from off-the-shelf equipment. Results showed that CABP was more correlated with the proposed methodology in comparison to all PBP variables assessed; confirming our hypothesis that the CABP is the most suitable parameter to collate through the time elapsed from ECG R-wave to the BCG J-wave. That is, considered featured time (RJ-interval) includes the time of a pulse pressure propagating at an aortic district. Bland-Altman plots showed an almost zero mean error (\u\ < 0.02mmHg) and bounded standard deviation o < 5mmHg for all systolic and mean central BP readings. Pilot work provided a landmark in order to develop a compact device that allows the integration of wireless blood pressure monitoring into a wearable system. Another contribution of this thesis is the proposition of a wearable device for PTT-computing by also including design considerations for the signal conditioning chains for ECG and BCG signals. The proposed design procedure takes care of minimizing the impact of spurious delays between physiological signals, which eventually degrade the PTT computation. Further, such a procedure could be suitable for any PTT-acquisition. Filtering with low and controlled delay is required for this biomedical application, and proposed conditioning chains provide less than 2ms group-delay, showing the effectiveness of the proposed approach. In order to provide the methodology with higher autonomy and integration, a highly miniaturized implementation of the filtering approach was also proposed. It includes the design of proposed architectures in CMOS technology to implement the particular low-delay filtering at reduced bandwidth featuring ultra-low-power characteristics. Results show that less than 2ms delay for the ECG QRS-complex can be achieved with a total current consumption of IDD = 2:1nA at VDD = 1:2V of power supply. Such development meant another significant contribution of this work in the conception of highly autonomous wearable devices for PTT acquisition. The first stage of validations on the wearable CABP estimation showed that, when considering data from one volunteer, results achieved with off-the-shelf equipment could be replicated by using a proposed wearable device, and the method could be further validated by using the wearable version. Additionally, CABP estimation from the proposed wearable device could be feasible by using three feature times (FTs) as CABP surrogates; that is, RI, RJ, and IJ intervals (from ECG and BCG wearable recordings). The first validation of the method also showed that CABP could be accurately predicted by the proposed methodology when in the order of daily calibrations are performed. The second stage of validations involved a study with a group of volunteers, and new alternatives were explored (twentyseven: nine PTTBMs along the three FTs) for the CABP estimation. We found that CABP could be accurately estimated (inside AAMI requirements) through the presented methodology by using four of the explored alternatives, whereas the RI interval, an FT lacking any PTT assessment, emerged as the best surrogate for the CABP estimation. Hence, a principle different from the traditional PTT-based method arises as a more advantageous method for the CABP estimation in the light of evidence reported in this validation, and, to our knowledge, this is the first time that CABP has been successfully estimated from a wearable device. The final significant contribution of this thesis meant the last chain-link in the process to achieve an utterly original method to estimate CABP. A novel PTTBM to estimate CABP is proposed, which uses a ow-driven two-element Windkesel network constructed from FTs extracted from the wearable recordings. When classic PTTBMs are applied, the fitting of parameters often leads to values without a physiological basis. Opposite to that in the proposed PTTBM, the parameters have a clear physiological meaning, and the parameter fitting led to values that are consistent with this meaning and more stable throughout calibrations. In conclusion, this thesis introduces a novel device that exploits an alternative and indirect method for CABP estimation. Variants of the principle used, accordingly, PTT method, have been previously explored to estimate PBP but not for central aortic BP. Additionally, the device was designed to be wearable; that is, it is attached to the clothes, causing low discomfort for the user during the measurement, thus, allowing continuous and ambulatory monitoring of aortic pressure. The developed wearable system, validated in a series of volunteers, showed promising results towards the continuous CABP monitoring. Se estima que casi un tercio de la población adulta mundial sufre de algún grado de hipertensión, siendo esto un factor de riesgo significativo para la enfermedad cardiovascular. La presión arterial (PA) es el parámetro utilizado para evaluar estos posibles estados de hipertensión; actualmente existe una necesidad de generación de nuevas tecnologías que permitan aumentar la frecuencia entre medidas de PA, pero al mismo tiempo de reducir el nivel de oclusión de éstas (técnicas aceptadas están mayoritariamente basadas en la oclusión y son de acceso limitado). El modelo Moens-Korteweg podría proveer los argumentos para la creación de nuevas técnicas para estimar la PA prescindiendo de cualquier brazalete inflable. Más específicamente, podría obtenerse una estimación indirecta de la PA a través de la medición del tiempo que tarda el pulso de presión en propagarse entre dos puntos vasculares predefinidos, método conocido como tiempo de tránsito del pulso (PTT). En la presente tesis se desarrolló un dispositivo vestible que explota este método alternativo e indirecto para la estimación de la PA pero a nivel central, es decir, busca estimar la PA en la aorta (CABP), la principal arteria de la red vascular. Para ello, los principales desarrollos de este trabajo incluyeron : prueba de concepto del método propuesto basado en PTT para estimar CABP, el desarrollo de un dispositivo vestible (incluyendo dos etapas de validaciones para la estimación de la PA), la propuesta de un circuito integrado para el hardware vestible y el desarrollo de un nuevo modelo para la estimación de la PA (PTTBM, es decir, la relación matemática que vincula las variables medidas con el hardware diseñado y la estimación de la PA). A continuación se presentan las principales contribuciones resultantes de cada frente de trabajo. Una de las contribuciones de esta tesis es el uso del principio PTT para estimar CABP en lugar de la BP periférica (PBP) (como se usa típicamente en la literatura). La prueba de concepto mostró la viabilidad de la estimación de CABP a partir del principio PTT mediante la adquisición de señales electrocardiograma (ECG) y balistocardiograma (BCG) utilizando equipos comerciales. Los resultados mostraron que CABP estaba más correlacionado con la metodología propuesta en comparación con todas las variables de PBP evaluadas; confirmando nuestra hipótesis de que la CABP sería la variable más adecuada para estimar a partir del tiempo transcurrido desde la onda R del ECG hasta la onda J del BCG. Es decir, el tiempo considerado (intervalo RJ) incluye un tiempo de propagación del pulso de presión a través de un segmento aórtico. Las gráficas de Bland-Altman mostraron un error medio casi nulo (\u\ < 0.02mmHg) y una precisión o < 5mmHg para las variables de presión sistólica y media centrales. La prueba de concepto proporcionó un hito para desarrollar un dispositivo vestible apuntando a la monitorización inalámbrica de la presión arterial en un sistema imperceptible para el usuario. Otra contribución de esta tesis es la propuesta de este dispositivo vestible para la adquisición de la PTT. El desarrollo incluye consideraciones de instrumentación necesarias para el correcto acondicionamiento de las señales ECG y BCG, de las cuales se obtiene la PTT. En particular, el procedimiento de diseño propuesto busca minimizar el impacto de los retrasos espurios entre las señales fisiológicas, que eventualmente degradan la computación de la PTT. Además, dicho procedimiento podría ser aprovechado por otros desarrolladores del método sin importar las definiciones de PTT que éstos usen. La limitación de banda con bajo retardo es necesario para esta aplicación biomédica, y el hardware de acondicionamiento propuesto proporciona menos de 2 ms de retraso en las se~nales (ECG y BCG) mientras consigue limitar sus bandas a decenas de Hz, lo que muestra la efectividad de la metodología propuesta. Adicionalmente, con el fin de proporcionar a la metodología de una mayor autonomía e integración, se propone una implementación altamente miniaturizada de la sección de filtrado con bajo retraso. Se incluye el diseño de nuevas topologías propuestas en tecnología CMOS para implementar el particular filtro de bajo retraso con reducido ancho de banda, y con características de ultra bajo consumo de potencia. El diseño integrado consigue obtener resultados similares al obtenido anteriormente (con componentes discretos) alcanzando un retraso de menos de 2 ms para el complejo QRS del ECG, pero con un consumo de IDD = 2:1 nA a un VDD = 1:2 V . Tal desarrollo significó otra contribución de este trabajo en el área de circuitos altamente autónomos para instrumentación biomédica. La primera etapa de validaciones en la estimación vestible de la CABP se basó en experimentaciones con un voluntario, mostrando que, la estimación vestible podría alcanzar los mismos resultados que los alcanzados utilizando equipos de investigación, permitiendo así avanzar en la validación del método propuesto utilizando el equipamiento vestible diseñado. Además de esto, se encontró que la estimación de CABP a partir del dispositivo vestible podría ser factible utilizando varios tiempos característicos (FT) extraídos de las señales vestibles ECG y BCG (intervalos RI, RJ e IJ) junto con un popular PTTBM. La primera validación del método también arrojó que la metodología propuesta podría estimar con precisión la CABP cuando el tiempo entre calibraciones es del orden de un día. La segunda etapa de validación implicó un estudio con un grupo de voluntarios, nuevas alternativas se exploraron esta vez (veintisiete: nueve PTTBM con tres FT) para la estimación de CABP. Descubrimos que CABP podría estimarse con precisión (dentro de los requisitos de AAMI) a través de la metodología presentada mediante el uso de cuatro de las alternativas exploradas, mientras que el intervalo RI, siendo un FT que a priori no tiene ninguna vinculación con un PTT, surge como el mejor estimador de la CABP. Se concluye entonces, que un principio diferente del método tradicional basado en PTT podría ser más ventajoso para la estimación de CABP a la luz de la evidencia encontrada en esta validación y, adicionalmente, a nuestro entender, esta es la primera vez que CABP se estima con éxito a partir de un dispositivo vestible. La contribución final de esta tesis significó el último eslabón de la cadena en el proceso de lograr un método completamente original para estimar CABP de punta a punta. Se propone un nuevo PTTBM para estimar CABP, éste es basado en una red Windkesel de dos elementos bajo una excitación de flujo. Estos elementos del PTTBM son construidos a partir de cantidades extraídas a través de procesamiento de las señales vestibles ECG y BCG. Cuando se aplican los PTTBM clásicos, el ajuste de sus parámetros (en calibración) a menudo conducen a valores sin base fisiológica, mostrando a su vez, una dispersión en sus valores a lo largo de distintas calibraciones que podrían ser inaceptables en la práctica. En contraposición, los parámetros del PTTBM propuesto convergen a cantidades con significado fisiologico claro y estable a lo largo de las calibraciones. En conclusión, esta tesis presenta un dispositivo novedoso que explota un método alternativo e indirecto para la estimación de CABP. El método propuesto es basado en la metodología de PTT, que si bien ha sido previamente explotado para estimar PBP, no se ha dirigido éste hacia el monitoreo vestible de la PA aórtica central. En este marco se desarrolla un dispositivo vestible, causando baja molestia en el usuario durante las mediciones, lo que permitiría un monitoreo continuo y ambulatorio real de la presión aórtica central. El sistema vestible desarrollado, validado en una serie de voluntarios, ha mostrado resultados prometedores hacia el monitoreo continuo de CABP. |
dc.format.extent.es.fl_str_mv | 157 p. |
dc.format.mimetype.es.fl_str_mv | application/pdf |
dc.identifier.citation.es.fl_str_mv | Fierro, G. Wearable estimation of central aortic blood pressure [en línea].Tesis de doctorado. Montevideo : Udelar. FI. IIE., 2020. |
dc.identifier.issn.none.fl_str_mv | 1688-2784 |
dc.identifier.uri.none.fl_str_mv | https://hdl.handle.net/20.500.12008/25156 |
dc.language.iso.none.fl_str_mv | en eng |
dc.publisher.es.fl_str_mv | Udelar.FI. |
dc.rights.license.none.fl_str_mv | Licencia Creative Commons Atribución - No Comercial - Sin Derivadas (CC - By-NC-ND 4.0) |
dc.rights.none.fl_str_mv | info:eu-repo/semantics/openAccess |
dc.source.none.fl_str_mv | reponame:COLIBRI instname:Universidad de la República instacron:Universidad de la República |
dc.subject.es.fl_str_mv | Wearable devices Blood pressure instrumentation modeling |
dc.title.none.fl_str_mv | Wearable estimation of central aortic blood pressure. |
dc.type.es.fl_str_mv | Tesis de doctorado |
dc.type.none.fl_str_mv | info:eu-repo/semantics/doctoralThesis |
dc.type.version.none.fl_str_mv | info:eu-repo/semantics/acceptedVersion |
description | Arterial hypertension affects a third of the world's population and is a significant risk factor for cardiovascular disease. Blood pressure (BP) is one of the most relevant parameters used for monitoring of possible hypertension states in patients at risk of cardiovascular disease. Hence, there exists a need for new monitoring solutions, which allow to increase the frequency between BP assessments, but also allow to reduce the level of occlusion in the attempts. Moens-Korteweg equation is among the main principles to estimate BP by dispensing of any inflatable cuff. This principle might lead to an indirect estimation of BP by measuring the time it takes the pressure pulse to propagate between two pre-established vascular points, accordingly the pulse transit time (PTT) method. This thesis proposes a wearable PTT-based method to estimate central aortic BP (CABP) and, the main milestones of this work included: proof of concept of the proposed method (pilot work), the development of a wearable device (including two stages of validation), the proposition of a miniaturized version (integrated circuit) of the analog front-end of the wearable hardware, and, the development of a novel PTT-based model (PTTBM, i.e., the mathematical relationship between measured variables and estimated BP) suitable for the proposed wearable methodology to estimate BP. The main contributions found at each milestone are presented. One of the contributions of this thesis is the use of the PTT-principle for estimating CABP instead of the peripheral BP (PBP) (as typically used in the literature). The pilot work showed the feasibility of CABP estimation from the PTT principle by using electrocardiogram (ECG) and ballistocardiogram (BCG) recordings from off-the-shelf equipment. Results showed that CABP was more correlated with the proposed methodology in comparison to all PBP variables assessed; confirming our hypothesis that the CABP is the most suitable parameter to collate through the time elapsed from ECG R-wave to the BCG J-wave. That is, considered featured time (RJ-interval) includes the time of a pulse pressure propagating at an aortic district. Bland-Altman plots showed an almost zero mean error (\u\ < 0.02mmHg) and bounded standard deviation o < 5mmHg for all systolic and mean central BP readings. Pilot work provided a landmark in order to develop a compact device that allows the integration of wireless blood pressure monitoring into a wearable system. Another contribution of this thesis is the proposition of a wearable device for PTT-computing by also including design considerations for the signal conditioning chains for ECG and BCG signals. The proposed design procedure takes care of minimizing the impact of spurious delays between physiological signals, which eventually degrade the PTT computation. Further, such a procedure could be suitable for any PTT-acquisition. Filtering with low and controlled delay is required for this biomedical application, and proposed conditioning chains provide less than 2ms group-delay, showing the effectiveness of the proposed approach. In order to provide the methodology with higher autonomy and integration, a highly miniaturized implementation of the filtering approach was also proposed. It includes the design of proposed architectures in CMOS technology to implement the particular low-delay filtering at reduced bandwidth featuring ultra-low-power characteristics. Results show that less than 2ms delay for the ECG QRS-complex can be achieved with a total current consumption of IDD = 2:1nA at VDD = 1:2V of power supply. Such development meant another significant contribution of this work in the conception of highly autonomous wearable devices for PTT acquisition. The first stage of validations on the wearable CABP estimation showed that, when considering data from one volunteer, results achieved with off-the-shelf equipment could be replicated by using a proposed wearable device, and the method could be further validated by using the wearable version. Additionally, CABP estimation from the proposed wearable device could be feasible by using three feature times (FTs) as CABP surrogates; that is, RI, RJ, and IJ intervals (from ECG and BCG wearable recordings). The first validation of the method also showed that CABP could be accurately predicted by the proposed methodology when in the order of daily calibrations are performed. The second stage of validations involved a study with a group of volunteers, and new alternatives were explored (twentyseven: nine PTTBMs along the three FTs) for the CABP estimation. We found that CABP could be accurately estimated (inside AAMI requirements) through the presented methodology by using four of the explored alternatives, whereas the RI interval, an FT lacking any PTT assessment, emerged as the best surrogate for the CABP estimation. Hence, a principle different from the traditional PTT-based method arises as a more advantageous method for the CABP estimation in the light of evidence reported in this validation, and, to our knowledge, this is the first time that CABP has been successfully estimated from a wearable device. The final significant contribution of this thesis meant the last chain-link in the process to achieve an utterly original method to estimate CABP. A novel PTTBM to estimate CABP is proposed, which uses a ow-driven two-element Windkesel network constructed from FTs extracted from the wearable recordings. When classic PTTBMs are applied, the fitting of parameters often leads to values without a physiological basis. Opposite to that in the proposed PTTBM, the parameters have a clear physiological meaning, and the parameter fitting led to values that are consistent with this meaning and more stable throughout calibrations. In conclusion, this thesis introduces a novel device that exploits an alternative and indirect method for CABP estimation. Variants of the principle used, accordingly, PTT method, have been previously explored to estimate PBP but not for central aortic BP. Additionally, the device was designed to be wearable; that is, it is attached to the clothes, causing low discomfort for the user during the measurement, thus, allowing continuous and ambulatory monitoring of aortic pressure. The developed wearable system, validated in a series of volunteers, showed promising results towards the continuous CABP monitoring. |
eu_rights_str_mv | openAccess |
format | doctoralThesis |
id | COLIBRI_ef69c031603f047102184f492997f012 |
identifier_str_mv | Fierro, G. Wearable estimation of central aortic blood pressure [en línea].Tesis de doctorado. Montevideo : Udelar. FI. IIE., 2020. 1688-2784 |
instacron_str | Universidad de la República |
institution | Universidad de la República |
instname_str | Universidad de la República |
language | eng |
language_invalid_str_mv | en |
network_acronym_str | COLIBRI |
network_name_str | COLIBRI |
oai_identifier_str | oai:colibri.udelar.edu.uy:20.500.12008/25156 |
publishDate | 2020 |
reponame_str | COLIBRI |
repository.mail.fl_str_mv | mabel.seroubian@seciu.edu.uy |
repository.name.fl_str_mv | COLIBRI - Universidad de la República |
repository_id_str | 4771 |
rights_invalid_str_mv | Licencia Creative Commons Atribución - No Comercial - Sin Derivadas (CC - By-NC-ND 4.0) |
spelling | Fierro Germán, Universidad de la República (Uruguay). Facultad de Ingeniería.2020-09-04T15:12:12Z2020-09-04T15:12:12Z2020Fierro, G. Wearable estimation of central aortic blood pressure [en línea].Tesis de doctorado. Montevideo : Udelar. FI. IIE., 2020.1688-2784https://hdl.handle.net/20.500.12008/25156Arterial hypertension affects a third of the world's population and is a significant risk factor for cardiovascular disease. Blood pressure (BP) is one of the most relevant parameters used for monitoring of possible hypertension states in patients at risk of cardiovascular disease. Hence, there exists a need for new monitoring solutions, which allow to increase the frequency between BP assessments, but also allow to reduce the level of occlusion in the attempts. Moens-Korteweg equation is among the main principles to estimate BP by dispensing of any inflatable cuff. This principle might lead to an indirect estimation of BP by measuring the time it takes the pressure pulse to propagate between two pre-established vascular points, accordingly the pulse transit time (PTT) method. This thesis proposes a wearable PTT-based method to estimate central aortic BP (CABP) and, the main milestones of this work included: proof of concept of the proposed method (pilot work), the development of a wearable device (including two stages of validation), the proposition of a miniaturized version (integrated circuit) of the analog front-end of the wearable hardware, and, the development of a novel PTT-based model (PTTBM, i.e., the mathematical relationship between measured variables and estimated BP) suitable for the proposed wearable methodology to estimate BP. The main contributions found at each milestone are presented. One of the contributions of this thesis is the use of the PTT-principle for estimating CABP instead of the peripheral BP (PBP) (as typically used in the literature). The pilot work showed the feasibility of CABP estimation from the PTT principle by using electrocardiogram (ECG) and ballistocardiogram (BCG) recordings from off-the-shelf equipment. Results showed that CABP was more correlated with the proposed methodology in comparison to all PBP variables assessed; confirming our hypothesis that the CABP is the most suitable parameter to collate through the time elapsed from ECG R-wave to the BCG J-wave. That is, considered featured time (RJ-interval) includes the time of a pulse pressure propagating at an aortic district. Bland-Altman plots showed an almost zero mean error (\u\ < 0.02mmHg) and bounded standard deviation o < 5mmHg for all systolic and mean central BP readings. Pilot work provided a landmark in order to develop a compact device that allows the integration of wireless blood pressure monitoring into a wearable system. Another contribution of this thesis is the proposition of a wearable device for PTT-computing by also including design considerations for the signal conditioning chains for ECG and BCG signals. The proposed design procedure takes care of minimizing the impact of spurious delays between physiological signals, which eventually degrade the PTT computation. Further, such a procedure could be suitable for any PTT-acquisition. Filtering with low and controlled delay is required for this biomedical application, and proposed conditioning chains provide less than 2ms group-delay, showing the effectiveness of the proposed approach. In order to provide the methodology with higher autonomy and integration, a highly miniaturized implementation of the filtering approach was also proposed. It includes the design of proposed architectures in CMOS technology to implement the particular low-delay filtering at reduced bandwidth featuring ultra-low-power characteristics. Results show that less than 2ms delay for the ECG QRS-complex can be achieved with a total current consumption of IDD = 2:1nA at VDD = 1:2V of power supply. Such development meant another significant contribution of this work in the conception of highly autonomous wearable devices for PTT acquisition. The first stage of validations on the wearable CABP estimation showed that, when considering data from one volunteer, results achieved with off-the-shelf equipment could be replicated by using a proposed wearable device, and the method could be further validated by using the wearable version. Additionally, CABP estimation from the proposed wearable device could be feasible by using three feature times (FTs) as CABP surrogates; that is, RI, RJ, and IJ intervals (from ECG and BCG wearable recordings). The first validation of the method also showed that CABP could be accurately predicted by the proposed methodology when in the order of daily calibrations are performed. The second stage of validations involved a study with a group of volunteers, and new alternatives were explored (twentyseven: nine PTTBMs along the three FTs) for the CABP estimation. We found that CABP could be accurately estimated (inside AAMI requirements) through the presented methodology by using four of the explored alternatives, whereas the RI interval, an FT lacking any PTT assessment, emerged as the best surrogate for the CABP estimation. Hence, a principle different from the traditional PTT-based method arises as a more advantageous method for the CABP estimation in the light of evidence reported in this validation, and, to our knowledge, this is the first time that CABP has been successfully estimated from a wearable device. The final significant contribution of this thesis meant the last chain-link in the process to achieve an utterly original method to estimate CABP. A novel PTTBM to estimate CABP is proposed, which uses a ow-driven two-element Windkesel network constructed from FTs extracted from the wearable recordings. When classic PTTBMs are applied, the fitting of parameters often leads to values without a physiological basis. Opposite to that in the proposed PTTBM, the parameters have a clear physiological meaning, and the parameter fitting led to values that are consistent with this meaning and more stable throughout calibrations. In conclusion, this thesis introduces a novel device that exploits an alternative and indirect method for CABP estimation. Variants of the principle used, accordingly, PTT method, have been previously explored to estimate PBP but not for central aortic BP. Additionally, the device was designed to be wearable; that is, it is attached to the clothes, causing low discomfort for the user during the measurement, thus, allowing continuous and ambulatory monitoring of aortic pressure. The developed wearable system, validated in a series of volunteers, showed promising results towards the continuous CABP monitoring.Se estima que casi un tercio de la población adulta mundial sufre de algún grado de hipertensión, siendo esto un factor de riesgo significativo para la enfermedad cardiovascular. La presión arterial (PA) es el parámetro utilizado para evaluar estos posibles estados de hipertensión; actualmente existe una necesidad de generación de nuevas tecnologías que permitan aumentar la frecuencia entre medidas de PA, pero al mismo tiempo de reducir el nivel de oclusión de éstas (técnicas aceptadas están mayoritariamente basadas en la oclusión y son de acceso limitado). El modelo Moens-Korteweg podría proveer los argumentos para la creación de nuevas técnicas para estimar la PA prescindiendo de cualquier brazalete inflable. Más específicamente, podría obtenerse una estimación indirecta de la PA a través de la medición del tiempo que tarda el pulso de presión en propagarse entre dos puntos vasculares predefinidos, método conocido como tiempo de tránsito del pulso (PTT). En la presente tesis se desarrolló un dispositivo vestible que explota este método alternativo e indirecto para la estimación de la PA pero a nivel central, es decir, busca estimar la PA en la aorta (CABP), la principal arteria de la red vascular. Para ello, los principales desarrollos de este trabajo incluyeron : prueba de concepto del método propuesto basado en PTT para estimar CABP, el desarrollo de un dispositivo vestible (incluyendo dos etapas de validaciones para la estimación de la PA), la propuesta de un circuito integrado para el hardware vestible y el desarrollo de un nuevo modelo para la estimación de la PA (PTTBM, es decir, la relación matemática que vincula las variables medidas con el hardware diseñado y la estimación de la PA). A continuación se presentan las principales contribuciones resultantes de cada frente de trabajo. Una de las contribuciones de esta tesis es el uso del principio PTT para estimar CABP en lugar de la BP periférica (PBP) (como se usa típicamente en la literatura). La prueba de concepto mostró la viabilidad de la estimación de CABP a partir del principio PTT mediante la adquisición de señales electrocardiograma (ECG) y balistocardiograma (BCG) utilizando equipos comerciales. Los resultados mostraron que CABP estaba más correlacionado con la metodología propuesta en comparación con todas las variables de PBP evaluadas; confirmando nuestra hipótesis de que la CABP sería la variable más adecuada para estimar a partir del tiempo transcurrido desde la onda R del ECG hasta la onda J del BCG. Es decir, el tiempo considerado (intervalo RJ) incluye un tiempo de propagación del pulso de presión a través de un segmento aórtico. Las gráficas de Bland-Altman mostraron un error medio casi nulo (\u\ < 0.02mmHg) y una precisión o < 5mmHg para las variables de presión sistólica y media centrales. La prueba de concepto proporcionó un hito para desarrollar un dispositivo vestible apuntando a la monitorización inalámbrica de la presión arterial en un sistema imperceptible para el usuario. Otra contribución de esta tesis es la propuesta de este dispositivo vestible para la adquisición de la PTT. El desarrollo incluye consideraciones de instrumentación necesarias para el correcto acondicionamiento de las señales ECG y BCG, de las cuales se obtiene la PTT. En particular, el procedimiento de diseño propuesto busca minimizar el impacto de los retrasos espurios entre las señales fisiológicas, que eventualmente degradan la computación de la PTT. Además, dicho procedimiento podría ser aprovechado por otros desarrolladores del método sin importar las definiciones de PTT que éstos usen. La limitación de banda con bajo retardo es necesario para esta aplicación biomédica, y el hardware de acondicionamiento propuesto proporciona menos de 2 ms de retraso en las se~nales (ECG y BCG) mientras consigue limitar sus bandas a decenas de Hz, lo que muestra la efectividad de la metodología propuesta. Adicionalmente, con el fin de proporcionar a la metodología de una mayor autonomía e integración, se propone una implementación altamente miniaturizada de la sección de filtrado con bajo retraso. Se incluye el diseño de nuevas topologías propuestas en tecnología CMOS para implementar el particular filtro de bajo retraso con reducido ancho de banda, y con características de ultra bajo consumo de potencia. El diseño integrado consigue obtener resultados similares al obtenido anteriormente (con componentes discretos) alcanzando un retraso de menos de 2 ms para el complejo QRS del ECG, pero con un consumo de IDD = 2:1 nA a un VDD = 1:2 V . Tal desarrollo significó otra contribución de este trabajo en el área de circuitos altamente autónomos para instrumentación biomédica. La primera etapa de validaciones en la estimación vestible de la CABP se basó en experimentaciones con un voluntario, mostrando que, la estimación vestible podría alcanzar los mismos resultados que los alcanzados utilizando equipos de investigación, permitiendo así avanzar en la validación del método propuesto utilizando el equipamiento vestible diseñado. Además de esto, se encontró que la estimación de CABP a partir del dispositivo vestible podría ser factible utilizando varios tiempos característicos (FT) extraídos de las señales vestibles ECG y BCG (intervalos RI, RJ e IJ) junto con un popular PTTBM. La primera validación del método también arrojó que la metodología propuesta podría estimar con precisión la CABP cuando el tiempo entre calibraciones es del orden de un día. La segunda etapa de validación implicó un estudio con un grupo de voluntarios, nuevas alternativas se exploraron esta vez (veintisiete: nueve PTTBM con tres FT) para la estimación de CABP. Descubrimos que CABP podría estimarse con precisión (dentro de los requisitos de AAMI) a través de la metodología presentada mediante el uso de cuatro de las alternativas exploradas, mientras que el intervalo RI, siendo un FT que a priori no tiene ninguna vinculación con un PTT, surge como el mejor estimador de la CABP. Se concluye entonces, que un principio diferente del método tradicional basado en PTT podría ser más ventajoso para la estimación de CABP a la luz de la evidencia encontrada en esta validación y, adicionalmente, a nuestro entender, esta es la primera vez que CABP se estima con éxito a partir de un dispositivo vestible. La contribución final de esta tesis significó el último eslabón de la cadena en el proceso de lograr un método completamente original para estimar CABP de punta a punta. Se propone un nuevo PTTBM para estimar CABP, éste es basado en una red Windkesel de dos elementos bajo una excitación de flujo. Estos elementos del PTTBM son construidos a partir de cantidades extraídas a través de procesamiento de las señales vestibles ECG y BCG. Cuando se aplican los PTTBM clásicos, el ajuste de sus parámetros (en calibración) a menudo conducen a valores sin base fisiológica, mostrando a su vez, una dispersión en sus valores a lo largo de distintas calibraciones que podrían ser inaceptables en la práctica. En contraposición, los parámetros del PTTBM propuesto convergen a cantidades con significado fisiologico claro y estable a lo largo de las calibraciones. En conclusión, esta tesis presenta un dispositivo novedoso que explota un método alternativo e indirecto para la estimación de CABP. El método propuesto es basado en la metodología de PTT, que si bien ha sido previamente explotado para estimar PBP, no se ha dirigido éste hacia el monitoreo vestible de la PA aórtica central. En este marco se desarrolla un dispositivo vestible, causando baja molestia en el usuario durante las mediciones, lo que permitiría un monitoreo continuo y ambulatorio real de la presión aórtica central. El sistema vestible desarrollado, validado en una serie de voluntarios, ha mostrado resultados prometedores hacia el monitoreo continuo de CABP.Submitted by Ribeiro Jorge (jribeiro@fing.edu.uy) on 2020-08-28T20:04:38Z No. of bitstreams: 2 license_rdf: 23149 bytes, checksum: 1996b8461bc290aef6a27d78c67b6b52 (MD5) Fie20.pdf: 10167872 bytes, checksum: ec0cba14edfb7a3a945ca205978223d5 (MD5)Approved for entry into archive by Machado Jimena (jmachado@fing.edu.uy) on 2020-09-04T15:11:13Z (GMT) No. of bitstreams: 2 license_rdf: 23149 bytes, checksum: 1996b8461bc290aef6a27d78c67b6b52 (MD5) Fie20.pdf: 10167872 bytes, checksum: ec0cba14edfb7a3a945ca205978223d5 (MD5)Made available in DSpace by Luna Fabiana (fabiana.luna@fic.edu.uy) on 2020-09-04T15:12:12Z (GMT). No. of bitstreams: 2 license_rdf: 23149 bytes, checksum: 1996b8461bc290aef6a27d78c67b6b52 (MD5) Fie20.pdf: 10167872 bytes, checksum: ec0cba14edfb7a3a945ca205978223d5 (MD5) Previous issue date: 2020157 p.application/pdfenengUdelar.FI.Las obras depositadas en el Repositorio se rigen por la Ordenanza de los Derechos de la Propiedad Intelectual de la Universidad de la República.(Res. Nº 91 de C.D.C. de 8/III/1994 – D.O. 7/IV/1994) y por la Ordenanza del Repositorio Abierto de la Universidad de la República (Res. Nº 16 de C.D.C. de 07/10/2014)info:eu-repo/semantics/openAccessLicencia Creative Commons Atribución - No Comercial - Sin Derivadas (CC - By-NC-ND 4.0)Wearable devicesBlood pressure instrumentation modelingWearable estimation of central aortic blood pressure.Tesis de doctoradoinfo:eu-repo/semantics/doctoralThesisinfo:eu-repo/semantics/acceptedVersionreponame:COLIBRIinstname:Universidad de la Repúblicainstacron:Universidad de la RepúblicaFierro, GermánSilveira, FernandoArmentano, RicardoUniversidad de la República (Uruguay). Facultad de IngenieríaDoctor en Ingeniería EléctricaLICENSElicense.txtlicense.txttext/plain; charset=utf-84267http://localhost:8080/xmlui/bitstream/20.500.12008/25156/5/license.txt6429389a7df7277b72b7924fdc7d47a9MD55CC-LICENSElicense_urllicense_urltext/plain; charset=utf-850http://localhost:8080/xmlui/bitstream/20.500.12008/25156/2/license_urla006180e3f5b2ad0b88185d14284c0e0MD52license_textlicense_texttext/html; charset=utf-838687http://localhost:8080/xmlui/bitstream/20.500.12008/25156/3/license_textd77747f0b79dbc4c411d2260a3d95cd2MD53license_rdflicense_rdfapplication/rdf+xml; charset=utf-823149http://localhost:8080/xmlui/bitstream/20.500.12008/25156/4/license_rdf1996b8461bc290aef6a27d78c67b6b52MD54ORIGINALFie20.pdfFie20.pdfapplication/pdf10167872http://localhost:8080/xmlui/bitstream/20.500.12008/25156/1/Fie20.pdfec0cba14edfb7a3a945ca205978223d5MD5120.500.12008/251562020-09-04 12:12:12.154oai:colibri.udelar.edu.uy:20.500.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Universidadhttps://udelar.edu.uy/https://www.colibri.udelar.edu.uy/oai/requestmabel.seroubian@seciu.edu.uyUruguayopendoar:47712024-07-25T14:44:17.484568COLIBRI - Universidad de la Repúblicafalse |
spellingShingle | Wearable estimation of central aortic blood pressure. Fierro, Germán Wearable devices Blood pressure instrumentation modeling |
status_str | acceptedVersion |
title | Wearable estimation of central aortic blood pressure. |
title_full | Wearable estimation of central aortic blood pressure. |
title_fullStr | Wearable estimation of central aortic blood pressure. |
title_full_unstemmed | Wearable estimation of central aortic blood pressure. |
title_short | Wearable estimation of central aortic blood pressure. |
title_sort | Wearable estimation of central aortic blood pressure. |
topic | Wearable devices Blood pressure instrumentation modeling |
url | https://hdl.handle.net/20.500.12008/25156 |