High precision voltage ratio measurements using a 3458A multimeter
Resumen:
This paper shows that very high precision ratio measurements (uncertainty of 3 parts in 10⁶) can be performed, also at frequencies of 1 kHz, using the multimeter HP 3458A. It is proposed to slightly change known algorithms to control the instrument, and a new uncertainty calculation is presented.
2006 | |
Calibration Divider Voltmeter Uncertainty |
|
Inglés | |
Universidad de la República | |
COLIBRI | |
https://hdl.handle.net/20.500.12008/30116 | |
Acceso abierto | |
Licencia Creative Commons Atribución - No Comercial - Sin Derivadas (CC - By-NC-ND 4.0) |
Sumario: | This paper shows that very high precision ratio measurements (uncertainty of 3 parts in 10⁶) can be performed, also at frequencies of 1 kHz, using the multimeter HP 3458A. It is proposed to slightly change known algorithms to control the instrument, and a new uncertainty calculation is presented. |
---|